APPLICATION
-
Before start the test, take all precautionary measures for operator’s safety (electrical protection gloves, etc.)
-
Even if it is possible to insert / remove energized conductors from the jaws (symbol ), de-energise (if possible) the conductors under test
-
Connect clamp’s output terminal to the inputs of the instrument to be used as indicator
-
Select the 700A desired range on clamp. The yellow LED “ON” should be on. Verify that “Low Bat” red LED it’s off. If no, open the back battery cover and replace the batteries.
-
Verify the zero at display of instrument to be used as indicator. If no, press ZERO key on clamp before start measurement
-
Open the clamp and put the tested conductor in the center of the clamp jaws. No gap is allowed between the connections of clamp jaws. Always consider the correct connection of clamp
-
Read the current value in the indicator’s display